Mitutoyo Surftest Sv400 User Manual Full ^hot^ Direct
Maintain a stable room temperature between 20°C ± 2°C to minimize thermal expansion errors.
The diamond stylus is highly fragile. Always raise the Z-axis safely away from the workpiece before removing a part. When not in use, slide the protective plastic stylus cover back on. Troubleshooting Common Issues Probable Cause Corrective Action Noisy or inconsistent data graphs External floor vibrations or air currents.
The Surftest SV-400 series features a dedicated desktop drive unit paired with a powerful data processing unit (often a dedicated controller or a PC loaded with Mitutoyo’s Surfpak software). Key Technical Specifications
Without the manual, users often search forums for fixes. Here are three issues the manual resolves directly: mitutoyo surftest sv400 user manual full
: The system uses interchangeable nosepieces to support both modes.
The is the Toyota Hilux of surface measurement—indestructible, boringly accurate, and awkward to operate.
Always use the motorized joystick or manual handwheels provided. Forced manual movement can strip the internal precision lead screws. Maintain a stable room temperature between 20°C ±
The SV-400 can store up to 10 measurement conditions (Cond 0 to Cond 9). The manual provides a matrix table showing which parameters are active per condition. For example:
Even with regular maintenance, you might encounter occasional issues with the SV-400. Consulting the troubleshooting section of the official user manual is always the best first step. However, the following table outlines some common problems and potential solutions:
Export your SURFPAK environment settings, calibration histories, and custom part measurement macros to an external storage drive. Proper Storage When not in use, slide the protective plastic
[ Mount Calibration Block ] │ ▼ [ Level the System using DAT Screen ] │ ▼ [ Input Nominal Value from Specimen Label ] │ ▼ [ Execute Auto-Calibration ] │ ▼ Verify: Measured Value == Nominal Value? ├── Yes ──> [ Calibration Complete ] └── No ──> [ Check Stylus / Re-Level ] Step-by-Step Auto-Calibration Procedure Mitutoyo Surftest SV 400 - Wolverine Machine
Keep relative humidity between 30% and 80% non-condensing to prevent electronic corrosion.